wet gas flow meter
1.Main performances
Model XML-1 XML-2 XMF-1 XMF-2 Drum capacity(m3) 0.002(2L) 0.005(5L) Rated flow(m3/h) 0.2(200L) 0.5(500L) Top flow (m3) 100 100 Excess flow(m3/h) 0.3(300L) 0.75(750L) Min. scale value(m3) 0.8×10-5 0.2×10-4 Normal pressure (Pa) 1000 1000 Accuracy in rated flow % /-1 /-1 Pressure in rated flow Pa </= 80 80 Outline dimensions(L/W/H)(mm) 490×264×220 490×264×304 Net weight(Kg) 4.5 5.4
2.Measuring medium
XML type is normal-type, it adopts brass material, and generallyused in the scope of
non-corrosive gases.
XMF type is anticorrosive-type, it adopts stainless steel material,can measure corrosive gases.
3.Usage
1. Put the meter on the worktable, adjust screws to enable blisterof leveler in the centre, and
maintain for a long time in use.
2. Open sealed screw cap of water level controller, pull outinternal wool rope.
3. In the plug jack of thermometer or manometer, inject distilledwater into the meter, when
redundant distilled water drain from hole of water level controller( sometimes it is clean when
outflow one drop), then put wool rope in sealed screw cap of waterlevel controller, and fasten the
sealed screw cap.
4. Install thermometer and manometer, (every case 10Pa)
5. Connect air channel according to direction of gas outlet andintake, and sealled.
6. Open gas valve, and then you can measure gas.
4.Attentions while using
1. When you use it, you must always pay attention to the waterlevel in the meter, otherwise it
will affect measuring accuracy.
2. When you use it, temperature should be maintained at 15 ~ 25???, and the temperature is the
same as room temperature, its temperature difference should be ≤ 2???.
3. You’d better readout in a few weeks after the operation of themeter.
4. When pressure of measuring gas is go beyond normal pressure 1 ~6 times (1 ~ 6 KPa), the
meters can still work, at this time manometer should be removed ,then use rubber plug plugs up
mounting holes of manometer.
5. When flow of measuring gas is go beyond rated flow, the meterscan still work, but measuring
accuracy will be greatly decreased.
7. The meter should not be installed in a cold room, for fear thatit will freeze inside.
8. In normal circumstances, check out accuracy every year atleast.
5.Accessories
1 pc of thermometer
1 pc of manometer
2 rubber plugs
6. Diagram of the meter
accuracy: /-1%
accessory: thermometer,manometer,rubber plugs
color: green or yellow
model: corrosive type, normal type
- AdditionalInformation
Payment Terms T/T, Western Union, Paypal, Others Minimum Order 1pc Lead Time DeliveryTime 7 days Sample Available No FOB Port Shanghai Loading Info usually by express Packing Info carton Target Market(s) Africa, Eastern Europe, Hong Kong/Macao/ Taiwan, Japan & Korea, Latin America, Mainland China, Middle East, North America, Northern Europe, Oceania, South Asia, SoutheastAsia, Western Europe