1. WorkCondition
1.1 Ambient temperature: 5âÂ??Â??ï½Â??43âÂ??Â??
1.2 Relative humidity: ≤90%RH (Under without condensationstatus)
1.3 Power supply: AC 90Vï½Â??AC 242V, 50HZ
1.4 Built-in battery: DC 12V 1800mAh rechargeable Ni-MH battery,the battery can be used continuously for 2 hours after full charged(24hours) under 25ml/h flow rate.
1.5 Atmospheric pressure: 550hPaï½Â??1060hPa
1.6 Storage temperature: -40âÂ??Â??ï½Â?? 55âÂ??Â??
1.7 Storage humidity: ≤95%RH
2. Basicparameter
2.1 Outline size (mm):140ï¼Â??Lï¼Â??×150ï¼Â??Wï¼Â??×220ï¼Â??Hï¼Â??
2.2 Weight: about 1.8kg
2.3 Power consumption: ≤10VA
2.4 Flow rate options: 1ml/hï½Â??1200ml/h, in 1ml/hincrement
2.5 Infusion total amount setting: 1mlï½Â??9999ml, in 1mlincrement
2.6 Occlusion pressure level:
Highï¼Â?? 825mmHg ± 150 mmHg (110kPa±20kPa)
Middleï¼Â??600mmHg ± 225mmHg (80kPa±30kPa)
Lowï¼Â?? 450mmHg ± 150 mmHg (60kPa±20kPa)
2.7 Flow rate error: ±5%
2.8 Alarm: occlusion, finish, low battery, standby (nooperation), air in line, door open, motor troubles, AC powerdisconnected, starting up by battery
2.9 Function: self-test, infusion amount setting, flow rate andinfused volume display, power indication, safe flow rate limit,data preserves, KVO (Keep Vein Open) function