Specifications:
Conditions of power supply: Voltage A. C380V/220V± 10%
Frequency: 50Hz± 1Hz
Capacity: 20KVA
Power supply resistance:
A. C380V≤ 1.0 Ω ; A. C220V≤ 0.03Ω
Maximum tube Voltage: 125KV
Maximum tube current: 200mA
Fuloroscopy: Tube voltage: 40-110KV
Tube current: 0.5-5mA
Radiograph: Tube voltage: 40-125KV
Tube current: Small focal spot: 25mA, 50mA, 100mA; Big
Focal spot: 150mA. 200mA
Spotfilm: 100mA (small focal spot)
High voltage generator: Capacity 15KW
Type of X-ray tube: Model: XD51-20, 40/125
Focal spot: Spotsmall focal spot: 1x1mm, big
Focal spot: 2x2mm
Rotate speed of anode: 2800r/min
Diagnostic table: Scope of rotation: 90° ~0° ~-5°
Mobile bean filtration device Grid density: N=28
Focusing distance: Fo>10000mm Grid ratio: R=8
Distance from surface to film: ≤ 60mm
Stomach&intestine radiograph device: Spotfilm device specifications:
127x178mm(5"x7") Binary picture/ 203254mm(8"x10")Binary picture
/254x305mm(10"x12")Binary picture/ 280x356mm(11"x14")Binary picture
Moving lengthways
(along groud railway) 1500mm
Moving crosswise
(left and right) 740mm
Moving up and
Right: 700-1900mm
Way of down:
Electromagnetism locking
Beam limited device: Manually single automatic muiti-optional
Table for radiograph: 2000x760x680mm
Movable range of bed facee Vertical: ± 450mm
Horizontal: ± 110mm
Column longitudinal movement: 1000mm
Film from the bed: ≤ 60mm
Tube moves up and down: 1100mm-1760mm
Radiography with Gird for cassette
Specifications: 10'x12', 11'x14', 12'x15'
Features:
1, Double diagnostic bed and double X-ray tube.
2, Equipped with KV. MA. S chain protection device, time limited protection device
Of radiograph.
3, Equipped with the manostat for filament of X-ray tube and the spacecharge
Compensator as to ensure the stable and accurate output. There is no ceiling
Track and install conveniently, easy to operate.
4, The sport film device of screen can use various sizes film jakect.
5, Equipped with magnetic lock.
200mA Medical Diagnosis X-ray Equipment (JQ200-2)
2016-11-05 05:45 Views:116
Price:Negotiable
Send Inquiry
Contact Infomation
Send Inquiry
Company name:Hangzhou Jakee Technology Co., Ltd.
Status:[Offline] [Send message] [Chat]
Business contact:Eason WANG(Mr.)
Telphone:
Fax:
Area: Caribbean Region-Bermuda
Address:RM.1604 HUARONG TIMES MANSIOM,NO.3880 JIANGNAN AVENUE.BINJIANG DISTRICT,HANGZHOU
Status:[Offline] [Send message] [Chat]
Business contact:Eason WANG(Mr.)
Telphone:

Fax:

Area: Caribbean Region-Bermuda

Address:RM.1604 HUARONG TIMES MANSIOM,NO.3880 JIANGNAN AVENUE.BINJIANG DISTRICT,HANGZHOU