Purpose:
The instrument is designed and made as per the National Standard of China GB/T3536-2008 "Test Methods for Flash and Fire Point of Petroleum Products (Cleveland Open Cup Methods)", and the Company Standard of Shanghai Q/YXYY15 "SYD-3536 Cleveland Open Cup Flash Point Tester". It is suitable to determine flash point and fire point of petroleum products having a flash point higher than 79 º C, excepting fuel oils.
Main technical specification and parameters:
1. Power supply: AC 220V± 10%, 50 Hz;
2. Heating device: It is a quartz tube heater, so there is no flame appears. Explosion protective; Its power can be continuouslyadjusted in the range of ~600W;
3. Flame application device: Automatically scan;
4. Thermometer: -6~400º C; Each scale division is 2 º C; As per GB/T514;
5. Ignition device:
(1) Ignition source: Coal gas (or any other combustible gas);
(2) Spout aperture: 0.6~0.8 mm;
6. Ambient temperature: -10º C~50º C;
7. Relative humidity: ≤ 85%;
8. Total power consumption: Not more than 650 W;