Metallurgy, Mineral & Energy
Volt/mA Calibrator (YHS715)
2016-11-04 03:12  Views:297
Price:Negotiable
Send Inquiry

Volt/mA Calibrator YHS715

YHS715 Volt/mA Calibrator is a source and measurement tool. This Calibrator is use to measure or output 0 to 24 mA DC current loop, and 0 to 10 V DC voltage. But the calibrator cannot be used to measurement and source simultaneously.
YHS715 Volt/mA Calibrator include this accessories: Holster, a pair of Test Leads, 9V battery, and this manual.

Specification
Specification are based on a one year calibration cycle and apply from +18deg to +28deg unless stated otherwise. "Counts" means number of increments or decrements of the least significant digit.

DC V Input and Output
Range Resolution Accuracy ± (% of reading + Counts)
100 mV 0.01 mV 0.02 % + 2
10 V 0.001 V 0.02 % + 2
Input impedance 2MΩ (nominal), < 100pF
Over voltage protection 30 V
Voltage driver capability 1 mA

DC mA Input and Output
Range Resolution Accuracy ± (% of reading + Counts)
24 mA 0.001mA 0.015 % +4
Overload protection 125 mA, 250V fast acting fuse
Percent display 0%=4mA, 100%=20mA
Source mode compliance 1000Ω at 20mA for battery voltage ≥ 6.8V,
(700Ω at 20mA for battery voltage 5.8 to 6.8V)
Simulate mode External loop voltage requirement: 24V nominal, 30V maximum, 12V minimum.

LOOP POWER
24 V ± 10%

General Specifications
Maximum voltage applied between any jack and earth ground or between any tow jack: 30V
Storage temperature: -40deg -60deg
Operating temperature: -10deg-55deg
Operating altitude: 3000 meters maximum
Temperature coefficient: ± 0.005% of range per deg for the temperature range -10deg to 18deg and 28deg to 55deg
Relative humidity: 95% up to 30deg, 75% up to 40deg, 45% up to 50deg, 35% up to 55deg
Shock: Random 2g, 5Hz to 500Hz
Safety: 1 meter drop test
Power requirements: 9VDC, 006P or 1604A
Size: 190mm L × 89mm W × 42mm H
Weight: 350g (include battery)

Contact Infomation
Send Inquiry